VXi VT1422A User Manual

Page of 529
Acquisition
Acquisition
Technology
VXI
Acquisition
Online at vxitech.com 
949 • 955 • 1VXI
213
Algorithmic Closed Loop Controller and 
Remote Channel Multifunction
VT1415A/1422A
Timing Signals
 
Timing:
   
 
Scan-to-scan timing and    
 
 
 
 
sample-to-sample  
 
 
 
 
 
timing can be set  
 
 
 
 
 
independently.
Scan triggers:
 
 
Can be derived from a software  
 
 
 
 
command or a TTL level from 
 
 
 
other VXI modules, internal timer, 
 
 
 
or external hardware.
 
 
 
Typical latency 17.5 µs.
Synchronization:
   
Multiple VT1415A/VT1422A  
 
 
 
 
modules can be synchronized 
 
 
 
at the same rate using the TTL  
 
 
 
trigger output from one 
 
 
 
VT1415A/VT1422A to trigger 
 
 
 
the others.
Alternate  
 
Multiple VT1415A/VT1422A
synchronization:
   
modules can be synchronized at    
 
 
 
different integer-related rates 
 
 
 
using the ALG:SCAN:RATIO  
 
 
 
 
command and the TTL trigger 
 
 
 
output from one VT1415A/ 
 
 
 
VT1422A module to trigger 
 
 
 
the others.
Scan Triggers
 
Internal:
  
 
100 µs to 6.5536 s
Resolution:
 
 
100 µs
Trigger count:
 
 
1 to 65535 or infinite
Sample Timer 
 
VT1415A: 10 µs to 32768 ms
Range:   
 
VT1422A: 40 µs to 32768 ms
Resolution:
 
 
0.5  µs
Measurement Specifications
The following specifications include the SCP and scanning 
A/D performance together as a unit. Accuracy is stated for 
a single sample. Averaging multiple samples will improve 
accuracy by reducing noise of the signal. The basic VT1415A 
scanning A/D has a full-scale range of ±16 V and five auto-
ranging gains of x1, x4, x16, x64, and x256. An SCP must be 
used with each eight channel input block to provide input 
protection and signal conditioning.
Note: For field wiring, the use of shielded twisted pair 
wiring is highly recommended.
Measurement resolution:
  16 bits (including sign)
Maximum reading rate:
 
VT1415A: Up to 56 kSa/s   
 
 
 
dependent upon  configuration
 
 
 
VT1422A: Up to 25 kSa/s    
 
 
 
 
dependent upon configuration
Memory:
 
 
64 kSamples
Maximum input voltage:
  Normal mode plus  
 
 
 
 
 
common mode
 
Operating: 
 ±16 V peak
 
Damage level: 
 ±42 V peak
Maximum common mode voltage:
 
 
Operating: 
 ±16 V peak
 
Damage level: 
 ±42 V peak
SCP input impedance:
 
100 MΩ  differential
Maximum tare cal offset:
  62.5 mV range ±75% of full  
 
 
 
 
scale, other ranges ±25%    
 
 
 
 
of full-scale
 
Jitter:
 
Phase jitter scan-to-scan:
  80 ps rms
Phase jitter card-to-card:
  41 ns peak 12 ns rms
Measurement Accuracy
Typically ±0.01% of input level; varies with the SCP used. 
Specifications are 90 days, 23 °C ± 1 °C, with *CAL done after 
a 1hr warm-up and CAL:ZERO done within 5 minutes. Note: 
Beyond the 5min. limitation and CAL:ZERO not done, apply the 
following drift error: Drift = 10 µV/°C ÷ SCP gain, per °C change 
from CAL:ZERO temperature.
Accuracy Data
Measurement accuracy is dependent upon the SCP module used. 
Refer to the accuracy tables and graphs for the individual SCP to 
determine the overall measurement accuracy. 
Many definitions of accuracy are possible. Here we use single-
shot with 3 sigma noise. To calculate accuracy 
assuming 
temperature is held constant within ±1 °C of the temperature at 
calibration, the following formula applies:
Single Shot 3σ = ±(√((GainError)2 + (OffsetError)2+ (3σ noise)2))
Correcting for Temperature