VXi Computer Hardware VX1410A User Manual

Page of 360
STATus Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–85
STATus:QUEStionable:TEMPerature:CMAXimum?
Queries the current maximum measurable temperatures in the mainframe since
power on. The returned value is in 
°
C.
There is no command associated with this query.
 "  ! !"  "  !
<Location>
Query response
{AMBient | OUT0 . . . OUT12 | DELTa0 . . .
DELTa12}
AMBient – ambient input temperature
OUT0 . . . OUT12 – output temperature of
each slot
DELTa0 . . . DELTa12 – output temperature
minus input temperature of each slot
<NR1>
0
–224, “Illegal Parameter Value”
You attempted to use an illegal parameter value.
None
Query:
         
Response:

STATus:QUEStionable:TEMPerature:LEVel?
STATus:QUEStionable:TEMPerature:MAXimum?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands