VXi Computer Hardware VX1410A User Manual

Page of 360
STATus Subsystem
3–106
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:SEVent?
Returns the summary information on the power supply voltage events, tempera-
ture events, blower events, filter events, and VXI monitor events.
A one in a bit indicates the corresponding event has been asserted. Unused bits
always return a value of zero.  The query returns a value of zero at power on or
when you send the *CLS command. Reading events by this command does not
clear the events. Reading the specific event register will clear the corresponding
bits in the register. (That is, STAT:QUES:VOLT? will clear all of the voltage bits
in this register). Summary information for current, power, triggers and IRQ
reside in an overflow register. If bit 30 is set, the overflow register should be
queried for more information. See the STATus:S2EVent? query on page 3–108.
There is no command associated with this query.
  
Parameter
Query response
Not applicable
<NR1>
Event status word bits
Definition
D0
Blower 1
D1
Blower 2
D2
Filter
D3
Slot 0 Delta Temperature
D4
Slot 1 Delta Temperature
D5
Slot 2 Delta Temperature
D6
Slot 3 Delta Temperature
D7
Slot 4 Delta Temperature
D8
Slot 5 Delta Temperature
D9
Slot 6 Delta Temperature
D10
Slot 7 Delta Temperature
D11
Slot 8 Delta Temperature
D12
Slot 9 Delta Temperature
D13
Slot 10 Delta Temperature
D14
Slot 11 Delta Temperature
D15
Slot 12 Delta Temperature
Syntax
Parameters