Valco electronic Valve Oven User Manual

Page of 62
Chapter 3
Operation
Operating Valves Automatically
40
6. Press CLEAR three times to exit to the 
Run Event 
table.
NOTE
Pressing CLEAR within a run table entry will delete that entry. To use CLEAR to back out 
of the menu, you must first be out of the run table entries themselves. 
Adding a Gas Switching Valve Event to the Run Table
For switching valves, two run time entries are required—one to switch the valves 
on and one to switch the valves off.
1. On the TRACE GC Ultra keypad, press RUN TABLE.
2. Scroll to 
Add run time event
 and press ENTER.
3. Select 
Valve
 and press ENTER.
4. Scroll to the Switching valve that you want to add and press ENTER.
The 
Run time
 parameter is the time into the run when the valve switches to the 
new position. The 
Setpoint
 parameter is the position that the valve will go to at 
the selected time. The 
On
 value usually designates a clockwise switch and the 
Off
 
value designates a counterclockwise switch.
If you want the valve to switch back to its original position at the end of run, then 
do not need enter a second run table entry.
SELECT PARAM to Add
  Valve #1
Sampling
  Valve #2
Switching
<
  Valve #3
Sampling
RUN TIME EVENT
  Valve #2
Switching
  Run time
0.00
  Setpoint
On (Off)