Sanyo Welder MCO-19M(UVH) User Manual

Page of 78
75 
     
 
 
Name Multi-gas 
Incubator 
Models MCO-19M 
MCO-19M(UV) 
MCO-19M(UVH) 
Accessories 
3 trays, 3 sets of tray supports, 3 gas tubes, 1 humidifying pan, 
1 gas injection nozzle, 1 gas injection nozzle tube 
1 pair of stacking plates A and B, 2 protective stickers, 6 tube bands
 
Optional accessories 
H
2
O
2
 Generator (MCO-HP) 
H
2
O
2
 Decontamination Kit (MCO-HL) 
Standard equipment 
UV Lamp Add-on Kit 
(MCO-19UVS) 
Standard equipment 
H
2
O
2
 Decontamination Reagent (MCO-H2O2), 
Automatic CO
2
 Standard Gas Calibration Kit (MCO-SG), 
Gas Pressure Regulator (MCO-100L), 
Extra Trays (MCO-47ST, includes 2 tray supports), 
Half Trays (MCO-25ST), Automatic CO
2
 Cylinder Changeover System (MCO-21GC), 
Roller Base (MCO-18RB), UV lamp replacement kit (MCO-20UV) 
Stacking spacer (MCO-18SB, when stacking on the MCO-175),   
Stacking spacer (MCO-21SB, when stacking on the MCO-20AIC) 
Ethernet (LAN) Interface (MTR-L03), RS232C/RS485 Interface (MTR-480) 
SANYO Data acquisition software (MTR-5000) 
4-20mA Interface (MCO-420MA, USA only) 
  Note: Designs and specifications are subject to change without notice. 
 
 
   PERFORMANCE
 
 
Temperature control range 
Ambient temperature+5 to 50
o
C (ambient temperature: 5 to 35
o
C) 
Temperature distribution 
± 0.25
o
C (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load) 
Temperature variation 
± 0.1
o
C (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load) 
CO
2
 control range 
0 to 20% 
CO
2
 variation 
± 0.15% (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load) 
O
2
 control range 
1 to 18% and 22 to 80% 
O
2
 variation 
± 0.2% (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load) 
Chamber humidity 
95% ± 5% 
Applicable environment 
condition 
Temperature: 5 to 35
o
C, Humidity: 80% max. 
(The designed performance may not be obtained 
  If the ambient temperature is equal or less than 15
o
C) 
Noise level 
30 dB (A scale) 
Power consumption 
Max. 354 W 
Max. 335 W
 
Heat emission 
Max. 1,274 kJ/h 
Max. 1,206 kJ/h 
Rated voltage, frequency 
110 to 120 V, 60 Hz 
220 V, 60 Hz 
220 to 240 V, 50 Hz 
Amperage
 
Max. 3.1 A
 
Max. 1.5 A 
Max. 1.5 A
 
Note: 
The unit with CE mark complies with EC directives. 
      Based on our measuring method. 
Designs and specifications are subject to change without notice.