Teledyne Vacuum Gauge 2002 User Manual

Page of 54
page 40   Model 2002
The 0-10V output option board, provides voltage outputs proportional to the HPM-2002
pressure reading.  The first channel corresponds to the higher-pressure range (0-1024) Torr).
The second channel corrresponds to the lower pressure range (0-1000mTorr).  The equation
below gives the output voltage for a given pressure.
   
          V(channel 1) =  
   P(Torr)
                          100
   
          V(channel 2) =  
   P(mTorr)
                          100
Where:
 V (channel 1) is the voltgage between pins 1 and 3,
 V (channel 2) is the voltgage between pins 8 and 7.
P is the indicated pressure.