Probe Compensation
TDS 620A, 640A, & 644A User Manual
3Ć111
4.
If you need to change the input impedance, press Coupling (main). Then
toggle the side menu selection 
W
 to get the correct impedance.
5.
Press SHIFT ACQUIRE MENU 
 Mode (main) 
 Hi Res (side).
6.
Adjust the probe until you see a perfectly flat top square wave on the
display. Figure 3-58 shows where the adjustment is located.
Figure 3-58: P6139A Probe Adjustment
See 
Probe Accessories, on page 3-97.
See 
Probe Selection, on page 3-112.