Omega Webcam OM-USB-1208FS User Manual

Page of 30
OM-USB-1208FS User's Guide 
Functional Details 
21 
Synchronizing multiple units  
You can run up to two OM-USB-1208FS units on most computers. 
You can connect the SYNC pin of two OM-USB-1208FS units together in a master/slave configuration and 
acquire data from the analog inputs of both devices using one clock. When the SYNC pin is configured as an 
output, the internal A/D pacer clock is sent to the screw terminal. You can use this signal as a clock input to a 
second USB by connecting it to the SYNC pin of the second USB. 
When used as a clock input, the SYNC pin operates in one of two modes – Continuous or Gated.  
In the default Continuous mode, a OM-USB-1208FS ignores the first clock pulse in order to ensure adequate 
setup time. Use this mode if the unit is being paced from a continuous clock source, such as a generator.  
In the Gated mode, it is assumed that the clock signal will be held off for an adequate amount of time for setup 
to occur. No clock pulses are ignored. Use this mode if the OM-USB-1208FS is set up as a slave and the source 
of the external clock is another USB. 
The SYNC pin (pin 19) is set for pacer output by default. To set it to input requires a software command (see 
Figure 13 below). 
To synchronize a master OM-USB-1208FS with a slave OM-USB-1208FS and acquire data, follow the steps 
below. 
1.  Connect the SYNC pin of the master OM-USB-1208FS to the SYNC pin of the slave OM-USB-1208FS. 
2.  Run InstaCal. 
3.  From the 
PC Board List
 on the 
InstaCal
 main form, double-click on the OM-USB-1208FS you want to use 
as a slave. The 
Board Configuration
 dialog opens. 
4.  Select Gated from the 
Ext. Clock Type
 drop-down list. 
5.  Set the Universal Library 
EXTCLOCK
 option with 
cbAInScan()/AInScan
 for the slave OM-USB-1208FS 
to enable pacing from the master USB device.  
This InstaCal option does not affect internally paced acquisition. It only affects scans that use the 
EXTCLOCK
 
option. 
An example of a master/slave configuration is shown below. 
 
Figure 13. Configuring for synchronous data acquisition 
When you are operating one OM-USB-1208FS, do not set the 
EXTCLOCK
 option unless you are using an 
external clock for A/D pacing.