Page 96 
 
www.PressureSystems.com 
4.3.2. Span 
Calibration 
Summary 
 
Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a 
9116 Intelligent Pressure Scanner. It is assumed that all channels in the unit are of the same full 
scale pressure range. Optional commands are shown within brackets [ ]. 
 
 
Description 
Model 9116 Command 
Ensure that valves are in RUN/CAL mode (default) 
[w1200] 
…  normal data acquisition 
 
Perform Re-zero calibration 
See Section 4.2.2 
Place the module calibration manifolds into the CAL position 
if this is the desired pressure application input. The RUN 
position may be a better choice for modules with 
transducers having different ranges. 
[w0C01] for CAL 
 
[w0C00] for RUN 
Apply exact full scale pressure to appropriate module CAL 
and CAL REF inputs [or optionally to RUN inputs]. 
 
Delay for settling of pneumatic inputs 
 
Verify that measured data reads near expected full scale 
[rFFFF0] 
Instruct module to calculate new gain coefficients for all 16 
channels 
ZFFFF 
Place calibration manifold back into the RUN position 
w0C00 
Store new gain coefficients to transducer non-volatile 
memory 
w09 
… continue normal data acquisition