Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 104 
 
www.PressureSystems.com 
Table 5.1 provides a convenient cross reference summary of the components found in your 
Model 9116 Intelligent Pressure Scanner.  This may be used as a guide to identify the 
appropriate component replacement sections in this chapter. 
 
Table 5.1 
Component Cross Reference 
 
Component 
Section
PC-322 Main Board PCB Assembly 
5.1.3.3 
PC-323 PowerPC daughterboard PCB 
Assembly 
5.1.3.1 
PC-327 Analog PCB Assembly 
5.1.3.2 
Internal Pneumatic Calibration 
Manifold 
5.1.6 
Internal Solenoid Valves 
5.1.5 
Internal DH-200 Transducer 
5.1.4 
 
5.1.1  
  Common Maintenance 
 
Your Model 9116 Intelligent Pressure Scanner is designed for rugged use.  No special 
preventive maintenance is required, although periodic maintenance may be required to replace 
worn or damaged components.   Upgrades or modifications of module hardware or firmware 
may also be periodically required.  For users who wish to do their own maintenance and repairs, 
maintenance kits and replacement parts for each model may be purchased from the factory.   
 
 
 
 
 
 
 
All circuit boards are sensitive to electrostatic discharges.  Anti-static protection is required 
whenever the unit is open. 
 
When performing any type of maintenance of Model 9116 components, the following guidelines 
and precautions should always be followed: 
 
● 
Verify that the work area and technicians are properly grounded to prevent damage to  
 
electronic components due to electrostatic discharge. 
 
 
 
 
 
 
 
 
 
 
● 
Ensure that all electrical and pneumatic connections have been removed from the module. 
 
● 
Ensure that the work area is free of dust and other possible contaminants that may affect  
 
the high tolerance machined parts (and pneumatic seals, if model has an integral  
 manifold). 
 
● 
Care must be taken to prevent contaminants from reaching O-ring surfaces.  If O-ring  
 
surfaces require cleaning, use a lint-free applicator with acetone to remove dirt and lightly  
 
lubricate the O-ring surface with Krytox® provided in the maintenance kit. 
 
● 
Never use sharp objects to cut tubing from the bulged tubes.  The tiny scratches left on  
 
the tubes could cause leaks.