Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 106 
 
www.PressureSystems.com 
5.1.3.1  
PC-327 Analog Board 
 
The following procedures should be used for replacement of the PC-327 Analog  Board.  Use 
the tools and follow the general precautions described in Section 5.1.1
 
(1)  Disassemble the module as described in Section 5.1.2.  Carefully remove the wiring 
harness from connector P1 of the PC-327 board.  Note the orientation of the PC-327 
relative to the rest of the module to ensure the new PC-327 is installed in the same position. 
 
(2)  Remove the two (2) Phillips-head screws securing the PC-327 board to the DH200 
transducers.  Carefully disconnect the PC-327 board from the DH200s by slowly working 
the board off; starting at one end and moving down the length of the board.  It is important 
that the gold pins are not bent when removing the board.  
 
(3)  Replace the old PC-327 board with a new one by placing it loosely on top of the DH200s.  
Ensure the board end containing connector P1 is oriented the same as the board just 
removed.  Inspect and make sure that all the gold pins fit easily into the female end of the 
connector on the DH200 transducers.  Press the board down evenly until all pins are firmly 
seated. 
 
(4)  Install the two (2) Phillips-head screws to secure the PC-327 to the DH200s.  Be careful not 
to over-tighten.  Install the wiring harness to connector P1 of the PC-327, ensuring proper 
pin 1 location
.  (Pin 1 of the ribbon cable has a red stripe while pin one of P1 will 
contain a square solder pad on the PC-327.)
 
 
(5)  Carefully align the  gasket on the top plate, ensuring it is free of contaminants.  Re-install 
the module electronics into the extrusion case.  Ensure that the alignment posts in the 
module's bottom panel align with the PC-327 or PC-322 electronics support brackets when 
placing the top panel and electronics back into the housing.   
 
(6)  Replace the screws that secure the top panel to the scanner housing and tighten.  Do not 
over- tighten; 7-9 inch-pounds torque should be sufficient. 
 
(7)  Test your scanner to ensure proper operation. 
 
5.1.3.2  
  PC-322/323 Main Board/PowerPC Daughter Board Assembly  
 
The following procedures should be used for replacement of the PC-322/323 Main 
Board/PowerPC Daughter board assembly.   Use the tools and follow the general warnings 
already described in Section 5.1.1
 
(1)  Disassemble the module as described in Section 5.1.2
 
(2)  Carefully remove any attached wiring harnesses from connectors  P3 and P6 of the PC-
322/323 board assembly. This will require cutting one nylon tie-wrap attached to the center 
mounting bracket.  The wiring harness from P1 will be disconnected in the following step.