5.1.3.3  
Remove and Replace PC-323 (Daughter board) on PC-322 (Main PCB) 
 
(1) 
Remove the two (2) Phillips-head screws that hold the PC-323 onto the PC-322 board.   
 
(Figure 5.3a
 
 
 
 
 
 
 
(2) 
Gently rock the PC-323 board back and forth to loosen it and then lift straight up to  
 
remove it.  Place the old PC-323 in an electrostatically-protected bag for possible repair  
 
at the PSI factory. 
 
 
 
 
 
 
(3) 
Remove the new PC-323 board from its electrostatically-protected container.  Ensure the  
 
nylon spacer is in place on top of the mounting bar and over the threaded hole.  (See  
 above 
photograph, 
Figure 5.3b).  Align the two (2) 40-pin connectors and press the  
 
board into place. 
 
 
Remove Phillips-head screws 
(screw already removed) 
 
Figure 5.3a 
Removing PC-323 Daughter Board 
Nylon spacer 
Figure 5.3b 
PC-323 Daughter Board Removed froom PC-322