Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 121 
 
www.PressureSystems.com 
Activate the Windows
®
 control bar (left click the START icon).  Select the SETTINGS line 
followed by the CONTROL PANEL folder.  In the CONTROL PANEL folder, select the 
NETWORK icon. Once in the NETWORK setup, select the tab labeled CONFIGURATION.   
Scroll through the list of installed configuration protocols.  Select the one labeled ‘TCP/IP-
>xxxx
’ where xxxx will typically identify your Ethernet adapter card.  There may be other TCP/IP 
protocols listed for other items such as dial up adapters, these are not used for the 
NetScanner
 System Ethernet configuration.  If the TCP/IP protocol is not listed in the 
configuration menu, left click the ADD button.  Continue by selecting to add a PROTOCOL
Select MICROSOFT from the Manufacturers list.  Then select TCP/IP from the networks 
protocol list. 
 
Once in the proper TCP/IP protocol setup, select the ‘IP Address’ tab.  Click on the button to 
enable the field ‘Specify IP Address.’  Once selected, the fields for IP address and Subnet will 
be enabled.  In the IP address, enter a TCP/IP address for your host computer.  An IP address 
of 200.200.200.001 will work if the Model 9116 is using the factory default IP address.  If the 
leftmost fields of the Model 9116 module IP addresses are different than the factory default of 
200.20x.yyy.zzz then the leftmost fields of the host computer’s IP address must match the 
module’s leftmost IP address field.  In the Subnet field a value of 255.0.0.0 can be entered for 
most configurations. 
 
When these fields are entered, click the OK icons until Windows
®
 prompts you to restart your 
computer. (Windows
®
 95/98 only).  Once the computer has restarted, it should be capable of 
communications with the Ethernet Model 9116 module. 
 
6.1.2.3  
  Verifying Host TCP/IP Communications 
 
At this point, the Model 9116 module should be configured to obtain its IP address through 
either static (default) or dynamic IP addressing.  The module’s IP address must be assigned 
and known in order to proceed.  The host computer has also been configured for TCP/IP 
protocol and assigned an IP address compatible with the Model 9116 IP address.  A simple 
method to verify proper operation is through the ping utility.  This is a simple TCP/IP utility that 
is found in Windows
®
 95/98/2000/XP/NT as well as most other TCP/IP packages.  The ping 
utility simply sends a test packet to the specified IP address and waits for reply to be returned.  
Model 9116 Ethernet modules are programmed to reply to these ping requests.   
 
To run the ping utility from Windows
®
 95/98/2000/XP/NT, follow these steps.  Left click the 
Windows
®
 START button. Move the mouse pointer to ‘RUN’ and left click on it.  At the prompt 
type ‘ping xxx.xxx.xxx.xxx’ where xxx.xxx.xxx.xxx represents the IP address of the device to 
test. The IP address of an Ethernet Model 9116 module should be used.  A small DOS window 
will appear as the ping application executes.  The ping program will either report that a reply 
was received or that it failed to receive a reply. If the ping application reported receiving a reply, 
the host computer and the Model 9116 module are both properly configured for TCP/IP 
communications.   
 
If an error free ping reply was not received, rerun the ping application using the IP address of 
the host computer. This will verify if the TCP/IP protocol was properly configured on the host 
computer.  If a ping reply was not received, verify the TCP/IP installation steps for your host 
computer.  Also verify that the host computer is configured for the proper IP address and subnet 
mask.