Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 25 
 
www.PressureSystems.com 
All modules effectively defer the host computer’s decision of “which channels of data do I want”  
until that time when the host chooses to send read commands to actually retrieve the desired 
data from the latest “buffered copy” of the continuously scanned, averaged, and engineering-
unit-converted data.  
 
See Section 3.1.4.4 (Delivery of Acquired Data to Host) for more information. 
 
While scanning, all modules take multiple samples and average each channel.  The number of 
samples per scanned channel
 defaults to 8, but may be set to one (to disable averaging) or to 
any  power of 2 (1, 2, 4, 8, 16, 32) to change the degree of averaging (and its effect on 
maximum scan rate).  The Set Operating Options (‘w’) command may change this variable at 
any time. 
 
3.1.4.3  
  Calibration Adjustment of Offset/Gain Correction Coefficients 
 
All Model 9116 Intelligent Pressure Scanners have built in software commands (and pneumatic 
hardware) to perform a periodic zero and span calibration adjustment of attached pressure 
transducers.  Use of these periodic adjustments result in the highest possible data accuracy.  
The result of these calibrations are a new set of internal offset and gain coefficients.  These 
correction coefficients are over and above those factory-determined and unchanging thermal 
correction coefficients stored in each transducer's non-volatile memory.  The factory coefficients 
provide the basic engineering unit conversion capability, while also correcting for various non-
linear effects, including temperature effect compensation.  The offset and gain correction 
coefficients provide for fine linear fit adjustment of the factory calibration of each transducer.  If 
used properly, the periodic zero and span calibration adjustment should be the only 
calibration required to maintain specified performance through the life of the Intelligent 
Pressure Scanner.
 
 
It is generally necessary for the transducer to have real zero and span pressure (specified as 2 
or more values) applied when calibration adjustment is required.  These pressure values may 
be generated by secondary pressure standards, such as the model 903x calibrator module or by 
other external means provided by the customer (such as a dead weight calibrator).  For the 
more common zero-only calibration adjustment, zero differential pressures can typically be 
provided without the need for external pressure generators.  All Model 9116 modules have built-
in pneumatic inputs (CAL side inputs) and calibration manifolds required for directing the 
generated pressures to the various channels of the module(s) being calibrated.  Refer to 
Chapter 4 of this manual for detailed background and procedures for periodic calibration 
of the Intelligent Pressure Scanners.
  A summary of the commands used for calibration 
purposes is included below. 
 
The Calculate and Set Offsets (‘h’) command is executed only when a suitable “minimum” 
(e.g., zero)  pressure has been applied to all channels of the module.  The new offset 
coefficients that result from execution of this command are stored in the module’s volatile (or 
temporary) engineering-unit conversion database.  They are also returned to the host in the 
command’s response.  
 
The Calculate and Set Gains (‘Z’) command should be executed only when “full-scale” (or 
other suitable specified up-scale) pressure has been applied to the appropriate channels of a 
module.  The new gain coefficients that result from this command are stored in the module’s 
volatile (or temporary) engineering-unit conversion database.  They are also returned to the host 
in the command’s response.