Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 33 
 
www.PressureSystems.com 
Example: 
 
 
● 
Configure and start the Multi-Point Calibration function so that it affects only the first  
 
four (4) channels of the module.  Three (3) pressure calibration points will by supplied  
 
when we continue this function later (see example for ‘01’ sub-command below).  A  
 
linear (1st order) fit will be used to obtain a new set of offset and gain correction  
 
coefficients for these four (4) channels.  The maximum average sample count (32) is  
 
used to collect each calibration data point, so as to minimize any noise in the data  
 
samples.  The module’s data acquisition process is altered immediately to collect the  
 
increased number of averages. 
 
 
C 00 F 3 1 32” 
 
 Read 
response: 
 
 
A”