Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 34 
 
www.PressureSystems.com 
Command ‘C’— Sub-command Index 01: Collect Data for a Calibration Point 
 
This sub-command has two (2) additional required parameters. 
 
Command 
C 01 pnt pppp.pppp” 
 
‘C’  is the command letter. 
‘ 01’ is the sub-command index (ii) for Collect Data 
 pnt’ identifies a particular calibration point that will be supplied.  It must 
be between 1 and npts, where npts was a parameter of the previously 
executed Configure & Start (‘00’) sub-command. 
‘ pppp.pppp’ is the pressure value (in current EU) that is actually applied 
currently to the module’s transducers by a precision calibrator. 
 
NOTE: all parameters are separated by a space. 
 
Response 
pppp.pppp [pppp.pppp]...” 
 
The actual measured pressure values (in current EU) from each affected 
channel of the module (highest numbered specified channel first, lowest 
numbered specified channel last, as per the pppp bit map parameter of the 
Configure & Start (‘00’) sub-command. The decimal response datum 
format (format 0) is always used. 
 
Description:  This sub-command (to be executed two or more times) carries out the data  
  
 
collection 
phase of the Multi-Point Calibration function for exactly one (1)  
 
 
  calibration point (i.e., per parameter pnt).  Each execution applies a specified  
 
 
  pressure value; then collects, averages, and stores the data for that calibration  
 
 
  point.  It must be repeated until all pressure points, as specified by the npts  
 
 
  parameter of the original Configure & Start (‘00’) sub-command, are applied and  
 
 
  their data collected.  For each particular point, enter the sub-command after that  
  
 
point’s 
pressure value has been properly applied to the module, and  that value is  
 
 
  stable (unchanging).  Pressure may be applied to either the Cal or Run ports, as  
 
 
  necessary.  Use a Model 903x Calibrator or some other suitable precision  
 
 
  pressure source to generate the pressure.  
 
It is not necessary to enter the two or more calibration points in strict numerical  
order (i.e., 1, 2, ... npts).   Any convenient entry order is allowed, though each  
point’s actual pressure value must be correctly stated (with the pppp.pppp  
parameter) when executed.  Previously entered points may be reentered if it is  
necessary to account for hysteresis.  However, all the points specified by the  
pts parameter of the Configure & Start (‘00’) sub-command must be supplied  
before the final Calculate and Apply  (‘02’) sub-command can be executed, else  
an error will result.