Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 35 
 
www.PressureSystems.com 
Example:  
● 
Supply each of the previously-specified three (3) pressure calibration points to the Multi- 
 
Point Calibration function, as was stated in the previous example of the Configure and  
 Start (‘00’) sub-command.  Assume that all the affected four (4) channels have  
 
differential transducers with the same -5 to +5 psi range.  Include at least one pressure  
 
point in the negative range of these transducers 
 
 
 
C 01 1 -2.5” 
 
C 01 2 0.0” 
 
C 01 3 5.0” 
 
 
Read responses (separately after each command executed  above): 
 
 
“-2.4998  -2.4999  -2.5001  -2.500” 
 
“0.0   0.0013   -0.0133   -0.00001” 
 
“5.0091   4.9992   5.0010  4.9998” 
 
 
 
Data are returned in reverse channel number order (i.e., 4, 3, 2, 1) in each response.