Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 37 
 
www.PressureSystems.com 
Example: 
 
● Finish 
the 
Multi-Point Calibration function previously started (as indicated by  
the previous examples of ‘C’ sub-commands ‘00’ and ‘01’).  Calculate new  
adjustment coefficients, and save them in the non-volatile memory of the  
module’s transducers.  These new coefficients will then be used for all  
subsequently calculated EU data acquired by the module, until another  
calibration function is performed in the future. 
 
 
 
 “C 02” 
 
 
 “w08” 
 
 
 “w09” 
 
 
 
Read responses (separately for each command executed above): 
 
 
 
 “A” 
 
 
 “A” 
 
 
 “A” 
 
If an error (“N”) response is returned on the first command, either the correct  
number of calibration points (per ‘00’ sub-command) were not supplied with  
reasonable pressure data values (via the multiple ‘01’ sub-commands), or the 
collected data yielded new calculated coefficients with unreasonable values.  In  
that case, the other two commands should not be used. 
 
 
 
  If execution of the last two ‘w’ commands is skipped above, the new calibration  
 
 
  data obtained will be stored only in volatile storage, and will be available for use  
 
 
  only until the module is RESET or loses power.