Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 54 
 
www.PressureSystems.com 
Command ‘c’— Sub-command Index 02: Stop Stream(s)
 
This sub-command is used to stop (or temporarily suspend) the delivery of any previously 
started host stream in a module, one at a time or all together, whether the stream was 
“continuous” or “limited.”  The sub-command’s format is: 
 
 
Command 
c 02 st” 
 
‘c’  is the command letter 
‘02’ is the sub-command index (‘ii’) for Stop Stream 
st’ is the stream id digit (single stream 1, 2, or 3, or 0=all streams)  
 
NOTE: all parameters are separated by a space
 
Response 
A” 
‘A’ is the acknowledge letter 
 
Autonomous 
Packet 
command stops generation of autonomous packets from the requested 
stream(s). 
 
 
Description:  This sub-command stops the current “run” of a particular specified host stream  
 
 
  (st=1-3), or stops the current  “run” of  “all configured” host streams with a single  
  
 
command 
(st=0).  
 
Any stopped stream may be resumed (i.e., restarted) with the Start Stream sub-
command as long as that stream remains defined in the module and any limited 
sequence count has not yet expired.  The Clear Stream sub-command may be 
used to undefine a stream.  Any continuous stream or unexpired limited stream 
that is restarted continues generating new sequence numbers (i.e., at the count 
where it left off when stopped).  However, the stream must be reconfigured with 
the Configure a Host Delivery Stream sub-command (00) before it restarts with 
sequence count =1.  Any expired limited stream must be reconfigured to restart 
at all.  
Example: 
● 
 
Stop all the streams configured in the previous example: 
 
 
 
 
c 02 0”  
 
  Read 
response:  
 
 “A”