Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 84 
 
www.PressureSystems.com 
 
ii 
dd 
Description 
‘q’ read 
index 
0C 
00 
 
01 
Set Cal Valves to RUN or LEAK Position (default)  
— choice made by ii=12. 
Set Cal Valves to CAL/RE-ZERO or PURGE Position — 
choice made by  ii=12. 
0B 
see chart
below 
0D-0E 
-- 
Reserved for factory use 
 
0F 
00 
01 
Disable periodic Thermal Coefficient Update task. 
Enable periodic Thermal Coefficient Update task (default). 
 
10 
01-20 Set 
Number of A/D Samples to Average,  (default = 8). 
Valid values are 4, 8,16, 32, and 64. Other values below 64 
are rounded up to the next valid value listed above. 
05 
11 
-- 
Reserved for factory use 
 
12 
00 
 
01 
Set Cal Valves to RUN or CAL/RE-ZERO Position (default) 
— choice made by ii=0C. 
Set Cal Valves to PURGE or LEAK Position — choice made 
by ii=0C. 
0B 
see chart
below 
13 
00 
01 
Use Static IP Address Resolution (default) 
Use Dynamic IP Address Resolution (RARP/BOOTP) 
(Results in immediately becoming the module’s new power-on 
default) 
06 
14 
00 
01 
 
 
02 
 
Disable Host Response/Stream Back-Off Delay (default). 
Enable Host Response/Stream Back-Off Delay as low-order 
byte of Ethernet Address(converted to decimal value * 20 
sec.). 
Enable Host Response/Stream Back-Off Delay specified 
per eeee as decimal value (* 20  sec.) 
07 
16 
00 
 
01 
Disable Host Response/Stream Total Size Prefix (default). 
Enable Host Response/Stream Total Size Prefix  (2-byte 
big-endian binary value with total size of response or stream 
data in bytes that follows it). 
08 
17 
00 Set 
TCP Connect Port per eeee as decimal value 
(default=9000). 
09 
18 
00 
01 
Disable Auto UDP Broadcast at Reset (default). 
Enable Auto UDP Broadcast at Reset
0A 
19 
00 
 
01 
Set Minimum Temperature Alarm Set Point (in 
N
C) 
per eeee as decimal value  (default = 0 
N
C). 
Set Maximum Temperature Alarm Set Point ( in 
N
C) 
per eeee as decimal value (default = 60 
N
C). 
0D 
 
0E 
1B 
00 Set 
Thermal Update Scan Interval per eeee as decimal 
value (seconds), 1 <= eeee <= 3600 seconds (default = 15). 
11