Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 86 
 
www.PressureSystems.com 
The Valve Position indexes (ii=12 and ii=0C) each have two states (00/01) that when combined 
provide four (4) possible states of the C1/C2/C3/C4 internal valves.  This “Logical Rotary 
Switch” with four (4) positions is summarized in the following chart: 
 
 
C1 Energized 
C2 Not 
(0C=01) 
C2 Energized 
C1 Not 
(0C=00) 
C3 Energized 
C4 Not 
(12=01) 
PURGE 
position 
LEAK/CHECK 
position 
C4 Energized 
C3 Not 
(12=00) 
CAL/RE-ZERO 
position 
RUN 
position 
 
 
Example: 
● 
 
Send TCP/IP commands to Model 9116 module (via its connected socket)  
 
setting the calibration valve to the CAL (or Re-Zero) position: 
 
 
 
w1200"   (Set RUN/CAL valve position) 
 
 
w0C01”   (Set 
CAL position) 
 
  
 
Responses 
(both 
commands): 
 
 
 
“A” 
 
 
“A” 
 
 
Note
 
If the programmer “knows” that the module is already in the RUN/CAL 
valve mode, the first command above is optional. For more information, 
see Figures 4.1 through 4.4 in Chapter 4