Measurement Specialties 9116 User Manual
Pressure Systems, Inc. Model 9116 User’s Manual
Page 86
www.PressureSystems.com
The Valve Position indexes (ii=12 and ii=0C) each have two states (00/01) that when combined
provide four (4) possible states of the C1/C2/C3/C4 internal valves. This “Logical Rotary
Switch” with four (4) positions is summarized in the following chart:
provide four (4) possible states of the C1/C2/C3/C4 internal valves. This “Logical Rotary
Switch” with four (4) positions is summarized in the following chart:
C1 Energized
C2 Not
(0C=01)
C2 Energized
C1 Not
(0C=00)
C3 Energized
C4 Not
(12=01)
PURGE
position
LEAK/CHECK
position
C4 Energized
C3 Not
(12=00)
CAL/RE-ZERO
position
RUN
position
Example:
●
Send TCP/IP commands to Model 9116 module (via its connected socket)
setting the calibration valve to the CAL (or Re-Zero) position:
“w1200" (Set RUN/CAL valve position)
“w0C01” (Set
CAL position)
Responses
(both
commands):
“A”
“A”
Note
If the programmer “knows” that the module is already in the RUN/CAL
valve mode, the first command above is optional. For more information,
see Figures 4.1 through 4.4 in Chapter 4.
valve mode, the first command above is optional. For more information,
see Figures 4.1 through 4.4 in Chapter 4.