HP E1368A User Manual

Page of 86
Parameters
Parameter Types.  The following table contains explanations and examples
of parameter types you might see later in this chapter.
Parameter Type
Explanations and Examples
Numeric
Accepts all commonly used decimal representations of
numbers including optional signs, decimal points, and
scientific notation.
123, 123E2, -123, -1.23E2, .123, 1.23E-2, 1.23000E-01.
Special cases include MIN, MAX, and DEF. MIN selects
minimum value available, MAX selects maximum value
available, and DEF selects default or reset value.
Boolean
Represents a single binary condition that is either true or
false.
1 or ON; 0 or OFF.
Discrete
Selects from a finite number of values. These parameters
use mnemonics to represent each valid setting.
An example is the TRIGger:SOURce <
source
> command
where 
source
 can be BUS, EXT, HOLD, or IMM.
Optional Parameters.  Parameters shown within square brackets ([ ]) are
optional parameters.  (Note that the brackets are not part of the command
and are not sent to the instrument.)  If you do not specify a value for an
optional parameter, the instrument chooses a default value.  For example,
consider the 
ARM:COUNt? [<MIN | MAX>]
 command.  If you send the
command without specifying a parameter, the present 
ARM:COUNt
 value is
returned.  If you send the 
MIN
 parameter, the command returns the
minimum count available (1).  If you send the 
MAX
 parameter, the
command returns the maximum count available (32,767).  Be sure to place a
space between the command and the parameter.
Linking Commands
Linking IEEE 488.2 Common Commands with SCPI Commands.  Use
a semicolon between the commands.  For example:
*RST;OUTP ON   
or
  TRIG:SOUR 
HOLD;*TRG
Linking Multiple SCPI Commands.  Use both a semicolon and a colon
between the commands.  For example:
ARM COUN 1;:TRIG:SOUR EXT
SCPI allows several commands within the same subsystem to be linked
with a semicolon.  For example:
ROUT:CLOS  (@100);:ROUT:OPEN  (A101)
or
ROUT:CLOS  (@100);OPEN  (@101)
Chapter 4
HP E1368A/69A/70A Command Reference   49