Hach sc1000 User Manual

Page of 150
97
Advanced Operations
6.3.3.5 Function set to WARNING working mode 
WARNING
WARNING LIST
Default value: Disabled
Sets the monitoring of the internal warning bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.
ERROR LIST
Default value: Disabled
Sets the monitoring of the internal error bits of the selected source. 
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.
PROCESS EVENT
Default value: Disabled
Sets the monitoring of the internal process event bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.
SET TRANSFER
Default value: DE-ENERGIZED
Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if some or all conditions (this means 
warning, error or process event bits) are detected in the selected source or if the source is 
missing. 
ON DELAY 
(0 s
999 s)
Default value: 5 seconds
Sets a delay time for the relay to turn on.
OFF DELAY 
(0 s
999 s)
Default value: 5 seconds
Sets a delay time for the relay to turn off.