Hach 900 MAX User Manual

Page of 160
Section 5
Page 71
8990options.fm
pH Probe
Figure 22
pH Probe Wiring to Junction Box (grounded)
Figure 23
pH Probe Wiring to Junction Box (un-grounded) 
5.2.2 pH Probe Programming
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
3.
Highlight pH using the 
UP
 and 
DOWN 
keys, then press 
SELECT
.
4.
Press the 
CHANGE CHOICE
 key to cycle between Logged and Not 
Logged, then press 
ACCEPT
.
5.
Enter a logging interval, then press 
ACCEPT
. Valid logging intervals are 
shown on the status bar along the bottom edge of the display. 
6.
Select another channel to configure or press 
RETURN
 to back up one 
step. Press 
MAIN MENU
 to return to the Main Menu display. 
5.2.3 pH Probe Calibration
Calibrate the pH probe after the pH probe is connected and programmed. 
Calibrating the pH probe requires a thermometer and any two of the following 
buffer solutions: 4, 7, or 10 pH. 
The pH probe is an application sensitive device. When used in harsh 
environments, the accuracy and life expectancy of pH probes can decrease. 