Hach DR 5000 User Manual

Page of 136
95
Advanced Operations
λ Setting wavelength
1. Press the 
λ key in the Options menu to select the wavelength 
range and the wavelength step.
2. Press the upper left key to open the numeric keypad and select 
the minimum wavelength. Press OK to confirm.
3. Press the upper right key to open the numeric keypad and 
select the maximum wavelength. Press OK to confirm.
Note: Do not select the same wavelength for minimum and maximum.
4. Highlight the required wavelength step. 
Note: Scan recordings of high resolution data take a longer time than 
recordings of low resolution data.Selecting a larger step allows the 
instrument to scan faster, but decreases the resolution of the collected 
data. 
Scale & Units
Scale: In the automatic scaling mode, the y-axis is automatically adjusted so that the total scan is 
displayed. 
The manual scaling mode allows sections of the scan to be displayed. 
Units: Choice of absorbance or transmittance.
Recall Data
Call up saved measurement data, wavelength scans or time courses, see 
.
Instrument Setup
Basic data of the instrument, see 
Table 14  Wavelength scan setup options (continued)
Option
Description