Setting the Time and Date
4.3 Setting the Time and Date
Before programming the sampler for the first time, check the real time clock 
and date settings. 
Note: When setting the time and 
date, the 
key causes the display to 
back up to the previously flashing 
character.
1.
Check the real-time clock and date settings by pressing 
TIME/READ
. The 
display will show the time and date for several seconds: 
10:35 AM 24OCT00
2.
If the time or date are not correct, press 
TIME/SET
. The time and date will 
appear on the display, with the hour flashing. 
3.
Press the appropriate numerical key(s) for the correct hour. When the 
correct hour flashes on the display, press 
YES
. This will cause the minutes 
to flash on the display. 
4.
Using the same procedure, select the correct minutes and press 
YES
. The 
am/pm indicator will flash. 
5.
If the desired indicator is flashing on the display, press 
YES
. If the desired 
indicator is not flashing, press 
NO
. This will cause the other indicator 
to flash. 
6.
Press 
YES
. The month will begin to flash. Select the correct day, month 
and year using the same procedure as you used to set the time.
7.
After the correct year has been entered, the display will read 
"Synchronize Time-enter at Time
"
. Pressing 
YES
 will start the clock. The 
display will read "Clock is Now Set
"
.
4.4 Bottle Setup Mode
When programming the sampler for the first time, you must first access the 
bottle setup mode which contains set-up information such as the units for 
bottle volume (milliliters and gallons) and the bottle volume. 
To access the Bottle Setup Mode, the sampler must be in standby state, the 
program cannot be in progress or "running". The sampler is in standby state 
when the display indicated any one of the following: "Ready to Start", 
"Program Complete", or "Program Halted". If the display reads "Program 
Running", press the 
CHANGE/HALT
 key. This will halt the program, thus 
placing the sensor in the “Program Halted” standby state. 
1.
With the sampler in standby state, press the * key. The display will read 
“Enter units for bottle volume, milliliters?”. 
2.
Press 
NO
 to cause other volume units to appear. When the desired units 
appear on the display, press 
YES
3.
The display will show 
"
Volume = _ _ _ _ _ mL
(or "Volume = _ _ _ _ _ 
Gallons
")
, whichever was chosen in the previous step. 
4.
Enter the volume for the individual sample bottle by pressing the 
numerical keys. 
5.
Press 
YES
 to accept the volume. 
This completes the bottle setup entries and the sampler is ready to program.