GE 100 User Manual

Page of 35
31
30
CAUTION
- Test results may vary if blood samples are taken from different sites or under certain 
conditions when glucose levels are changing rapidly such as: following a drink, a meal, an 
insulin dose or exercise. In these cases, only the fingertip should be used.
- DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood glucose).
- Fingertip samples can show rapid changes in glucose faster than palm or forearm samples.
- Consult  your healthcare professionals before sampling from your palm or forearm. 
- Since the blood flow in the forearm is less than that in the fingertips or palm, use the clear cap 
with the GE Lancing Device when testing on the forearm. 
- If you use the depth adjustable cap when testing on the forearm the blood sample may not 
large enough for the meter.
- Periodically compare the test system to a laboratory test system that is known to be well 
maintained and monitored by a healthcare provider.
Alternative Site Testing
View Window Appearance
Make sure  your  blood  sample  covers  the  whole 
area of the view window to get an accurate test 
result. An insufficient blood sample will result in 
an  error  message  ("Er4").  If  this  occurs,  repeat 
the test with a new test strip.
CAUTION
-  Check  the  expiration  date  printed  on  the  package  every  time  you  use  the  strip.  Do  not  use 
expired GE100 Blood Glucose Test Strips.
- Use each GE100 Blood Glucose Test Strip immediately after taking it  from the vial.
- Do not reuse GE100 Blood Glucose Test Strips.
- Do not perform testing immediately after moving from one ambient temperature to another 
(e.g.  after  coming  indoors    from  the  outside).    Allow  30  minutes  for  the  temperature  of  the 
meter and the test strips to equilibrate.
 Insufficient blood sample
 Enough blood sample