CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
Wetted materials may not be compatible
with process com 
position and operating
conditions. Application compat 
ibility is
entirely the responsibility of the user.
DANGER
HAZARDOUS AREA INSTALLATION
Installations near flammable liquids or in haz-
ardous area locations must be carefully eval-
uated by qualified on site safety personnel.
This device is not Intrinsically Safe or
Explosion Proof. 
To secure and maintain an intrinsically safe
installation, the certified safety barrier, trans-
mitter, and sensor combi 
nation must be
used. The installation system must comply
with the governing approval agency (FM,
CSA or BASEEFA/CENELEC) hazardous
area classification requirements. Consult
your analyzer/transmitter instruc tion manual
for details.
Proper installation, operation and servicing
of this device in a Hazardous Area Instal -
lation is entirely the responsibility of the user.
About This Document
This manual contains instructions for installation and operation of the Model TCL1056
Total Chlorine Analyzer.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
A
7/07
This is the initial release of the product manual. The manual 
has been reformatted to reflect the Emerson documentation 
style and updated to reflect any changes in the product offering.
B
11/07
Page 3 additions to 1.3 Specifications, page 5 changes to the 
Analyzer (option selection) table, page 14 updated the caution 
box and page 15 updated section 4.3.3.
C
7/10
Updated DNV logo.
D
5/11
Revised Sec. 1, specs 1.3, replaced fig 3-3, revised sec. 4.3.3
wiring, replaced fig 4.4. add to digital communication sec 8.0,
revise sec. 11.6.