Emerson 5300 User Manual
Manual Supplement
00809-0500-4530, Rev AA
October 2010
Rosemount 5300/5400 Series
1-10
ESTABLISH HART
COMMUNICATION
COMMUNICATION
The Rosemount 5300 Series and Rosemount 5400 Series can be configured
using the Rosemount Radar Master (RRM) PC software or a Field
Communicator. Configuration is done by sending HART commands through
the HART to Modbus Converter (HMC) to the 5300/5400 transmitter
electronics. To establish HART communication, connect to the MA/MB
terminals, or to the HART terminals. Both alternatives are described below.
using the Rosemount Radar Master (RRM) PC software or a Field
Communicator. Configuration is done by sending HART commands through
the HART to Modbus Converter (HMC) to the 5300/5400 transmitter
electronics. To establish HART communication, connect to the MA/MB
terminals, or to the HART terminals. Both alternatives are described below.
Connect to the MA/MB
terminals
terminals
The 5300/5400 level transmitter can be configured with RRM using the MA,
MB terminals.
MB terminals.
An RS-485 Converter is required to connect to the transmitter.
The transmitter will try to establish communication using different protocols
during 20 second timeslots from time of startup.
during 20 second timeslots from time of startup.
Figure 1-8. RS-485 Communication
after startup
after startup
The transmitter will continue to use a communication protocol once
communication has been established.
communication has been established.
HART
20 seconds
20 seconds
Modbus RTU
20 seconds
20 seconds
Configured
protocol
(Modbus RTU,
Levelmaster,
or Modbus
ASCII)
20 seconds
protocol
(Modbus RTU,
Levelmaster,
or Modbus
ASCII)
20 seconds
HART
20 seconds
20 seconds
Time
0 s
20 s
40 s
60 s
80 s
100 s
Configured
protocol
(Modbus RTU,
Levelmaster,
or Modbus
ASCII)
20 seconds
protocol
(Modbus RTU,
Levelmaster,
or Modbus
ASCII)
20 seconds