Cloning a Transmitter
19
Cloning a Transmitter
Cloning copies the Basic Setup information from one
transmitter to another. You can clone generic and
supported transmitters.
To use the calibrator to clone a transmitter:
1.  Connect the calibrator to the transmitter you want to
clone. You only need a communication connection.
2.  From the Active Device screen, press the 
Setup
softkey.
3. Use 
the 
u
 or 
d
 keys to select 
Basic
 from the list on
the screen and press 
e
.
 
 
pe09s.bmp
4. Press 
the 
Store Page
 softkey to load the settings into
the calibrator’s memory (not the transmitter).
5.  Disconnect the calibrator from the transmitter and
connect it to the transmitter you want to configure.
Press the 
Abort
 softkey to establish communication.
6.  Press the 
Setup
 softkey, then select 
Basic
 to return to
the Basic Setup screen.
7.  Press the 
Recall Page
 softkey to refresh the calibrator
display with the cloned parameter settings. The
settings are not transmitted to the transmitter at this
time.
8.  Use the 
u
 or 
d
 and 
e
keys to select
parameters one at a time, change them, or leave them
as cloned,  and press the 
Send
 softkey to transmit
them to the transmitter.