1-17. Wideband AC Voltage (Option 5700-03) Specifications
Table 1-21. Wideband AC Voltage (Option 5700-03) Specifications
Specifications apply to the end of the cable and 50 
Ω
 termination used for calibration
Range
Resolution
Absolute Uncertainty
±
5 °C from calibration temperature
30 Hz-500 kHz
Volts
dBm
24 Hours
90 Days
180 Days
1 Year
±
 (% output +
 µ
V)
1.1 mV
-46
10 nV
0.4 + 0.4
0
0.5 + 0.4
0
0.6 + 0.4
0.8 + 2
3 mV
-37
10 nV
0.4 + 1
0.45 + 1
0
0.5 + 1
0.7 + 3
11 mV
-26
100 nV
0.2 + 4
0.35 + 4
0
0.5 + 4
0.7 + 8
33 mV
-17
100 nV
0.2 + 10
0
0.3 + 10
0.45 + 10
0.6 + 16
110 mV
-6.2
1
 µ
V
0.2 + 40
0
0.3 + 40
0.45 + 40
0.6 + 40
330 mV
+3.4
1
 µ
V
0.2 + 100
2.5 + 100
0.35 + 100
0.5 + 100
1.1 V
+14
10
 µ
V
0.2 + 400
0.25 + 400
0.35 + 400
0.5 + 400
3.5 V
+24
10
 µ
V
0.15 + 500
0
0.2 + 500
0
0.3 + 500
0.4 + 500
Frequency
Frequency
Resolution
Amplitude Flatness, 1 kHz Reference
Voltage Range
Temperature
Coefficient
Settling
Time To
Full
Accuracy
Harmonic
Distortion
1.1 mV
3 mV
>3 mV
Hz
Hz
±
 (% output + floor indicated)
±
ppm/°C
Seconds
dB
00 
10 - 30
0.01
0.3
0.3
0.3
100
7
-40
00 
30 - 120
0.01
0.1
0.1
0.1
100
7
-40
120 - 1.2 k
0.1
0.1
0.1
0.1
100
5
-40
0
1.2 k - 12 k
1
0.1
0.1
0.1
100
5
-40
12 k - 120 k
10
0.1
0.1
0.1
100
5
-40
 120 k - 1.2 M
100
0
0.2 + 3
 µ
V
0
0.1 + 3
 µ
V
0.1 + 3
 µ
V
100
5
-40
 1.2 M - 2 M
100 k
0
0.2 + 3
 µ
V
0
0.1 + 3
 µ
V
0.1 + 3
 µ
V
100
0.5
-40
0  
2 M - 10 M
100 k
0
0.4 + 3
 µ
V
0
0.3 + 3
 µ
V
0.2 + 3
 µ
V
100
0.5
-40
  10 M - 20 M
1 M
0
0.6 + 3
 µ
V
0
0.5 + 3
 µ
V
0.4 + 3
 µ
V
150
0.5
-34
  20 M - 30 M
1 M
10.5 + 15
 µ
V
10.5 + 3
 µ
V
1 + 3
 µ
V
300
0.5
-34
Additional Operating Information:
dBm reference = 50 
Ω
Range boundaries are at voltage points, dBm levels are approximate.
dBm = 10 log 
(
Power
1 mW
)
 ; 0.22361 V across 50 
Ω
 = 1 mW or 0 dBm
Minimum output: 300
 µ
V (-57 dBm)
Frequency uncertainty: 
±
0.01 %
Frequency resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz.
Overload protection: A short circuit on the wideband output will not result in damage. After settling time, normal operation is
restored upon removal.