Fluke ScopeMeter 196 User Manual

Page of 116
Fluke 192/196/199
Users Manual
32
Displaying Recorded Data
In Normal view, the samples that roll off the screen are
stored in deep memory. When the memory is full,
recording continues by shifting the data in memory and
deleting the first samples out of memory.
In View All mode, the complete memory contents are
displayed on the screen.
4
Press to toggle between 
VIEW ALL
(overview of all recorded
samples) and
 NORMAL
 view.
You can analyze the recorded waveforms using the
Cursors and Zoom functions. See Chapter 4: “Using
Replay, Zoom and Cursors
”.
Using ScopeRecord in Single Sweep Mode
Use the recorder Single Sweep function to automatically
stop recording when the deep memory is full.
Continue from step 3 of the previous section:
4
Open the Recorder options
menu.
5
 
(2x)
Jump to the Mode field.
6
Select Single Sweep and accept
the recorder options.