Introduction and Specifications
Specifications
1
1-3
Table 1-2. Specifications (cont.)
Electromagnetic Compatibility
This instrument is designed to operate in Standard Laboratory
environments where in electromagnetic environment is highly
controlled. If used in areas with RF fields 
>
0.4 V/m there could be
errors in measurement.
DC Mode (including 5100 Series B)
Output Range
±
1 to 
±
19.9999 A
Accuracy of Output
±
(0.025 % of selected output + 1 mA)
Resolution
±
0.1 mA
Temperature Coefficient
±
(0.003 % of selected output + 100 
µ
A) / 
°
C, above 30 
°
C and
below 20 
°
C
Line Regulation
Output changes less than 0.001 % for a 
±
10 % change in line
voltage.
Load Regulation
Output changes less than 
±
(0.005 % + 0.3 mA) for a full load
changes of 4 V of compliance.
AC Mode (including 5100 Series B)
Output Range
1 A rms to 19.9999 A rms
Accuracy of Output
±
(0.07 % of selected output + 1 mA rms) from 50 Hz to 1 kHz,
and 
±
(0.07 % of selected output + 1 mA rms) x f from 1 kHz to 5
kHz, where f = frequency in kHz.
Resolution
±
0.1 mA rms
Temperature Coefficient
±
(0.003 % of selected output + 100 
µ
A rms) per 
°
C, above 30 
°
C
and below 20 
°
C.
Short Term Stability
Output changes less than 
±
(0.02 % + 500 
µ
A rms) in 10 minutes,
with constant line, load, and temperature.
Harmonic Distortion and Noise
±
(0.07 % of output + 1 mA rms) over frequency range of 30 Hz to
1 kHz a noise bandwidth of 300 kHz, 
±
(0.07 % of output + 1 mA)
x f from 1 kHz to 5 kHz, where f = frequency in Hz.
Line Regulation
Output changes less than 0.005 % for 
±
10 % changes in line
voltage.
Load Regulation
For frequencies less than 1 kHz:
±
(0.005 % lout + 150 
µ
A) per volt of output compliance voltage.
For frequencies greater than 1 kHz:
±
(0.005 % lout + 150 
µ
A) x f per volt of output compliance
voltage, where f = frequency in kHz.