Fluke IMPULSE 6000D User Manual

Page of 11
Impulse 6000D/7000DP Users 
Manual Supplement   
 
4/08 1 
Change #1, 
46470, 46697, 47714, 48007, 49680
 
Replace pages 31 through 35, with the following: 
Defibrillator Analyzer Specifications 
Energy Output Measurement 
Compatible Defibrillator Waveshapes
................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic 
Note 
AC Pulsed Biphasic waveform has not been approved in the United States. 
Autoranged Measurement
.................................... 0.1 to 600 J 
Accuracy 
0.1 to 360 J .....................................................
±(1 % of reading + 0.1 J) 
360 to 600 J ....................................................
±(1 % of reading + 0.1 J), typical 
Note 
For Pulsed Bi-Phasic defibrillator, specified accuracy is 
±
(1.5 % of reading + 0.3 J) on both ranges. 
Load resistance 
Resistance .......................................................... 50 
Ω  
Accuracy .........................................................
±1 %, non-inductive (<2 μH) 
Pulse trigger level
................................................. 20 V 
Pulse width 
Range ................................................................. 1.0 to 50.0 ms