Agilent Technologies Relay Matrix Switch Modules E1465A User Manual
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
Matrix Modules Command Reference 67
Chapter 4
Scanning Channels:
•
To scan a single channel use ROUT:SCAN (@ssrrcc)
•
To scan multiple channels use ROUT:SCAN (@ssrrcc,ssrrcc,...)
•
To scan sequential channels use ROUT:SCAN (@ssrrcc:ssrrcc)
•
To scan groups of sequential channels use ROUT:SCAN
(@ssrrcc:ssrrcc,ssrrcc:ssrrcc)
•
or any combination of the above
NOTE
Channel numbers can be in the <channel_list> in any random order.
Scanning Operation:
When a valid channel list is defined,
INITiate[:IMMediate] begins the scan and closes the first channel in the
<channel_list>. Successive triggers from the source specified by
TRIGger:SOURce advance the scan through the <channel list>. At the
end of the scan, the last trigger opens the last channel.
<channel_list>. Successive triggers from the source specified by
TRIGger:SOURce advance the scan through the <channel list>. At the
end of the scan, the last trigger opens the last channel.
Stopping Scan:
See ABORt
Related Commands:
TRIGger, TRIGger:SOURce
*RST Condition:
All channels open.
Example
Scanning Using External Device
See "Scanning Channels" in Chapter 3 for examples of scanning programs
using external instruments.
using external instruments.