                         Matrix Modules Command Reference   67
Chapter 4
Scanning Channels:
 
To scan a single channel use ROUT:SCAN (@ssrrcc)
To scan multiple channels use ROUT:SCAN (@ssrrcc,ssrrcc,...)
To scan sequential channels use ROUT:SCAN (@ssrrcc:ssrrcc)
To scan groups of sequential channels use ROUT:SCAN 
(@ssrrcc:ssrrcc,ssrrcc:ssrrcc)
or any combination of the above
NOTE
Channel numbers can be in the <channel_list> in any random order.
Scanning Operation:
 When a valid channel list is defined, 
INITiate[:IMMediate] begins the scan and closes the first channel in the 
<channel_list>. Successive triggers from the source specified by 
TRIGger:SOURce advance the scan through the <channel list>. At the 
end of the scan, the last trigger opens the last channel.
Stopping Scan:
 See ABORt
Related Commands:
 TRIGger, TRIGger:SOURce
*RST Condition:
 All channels open.
Example
Scanning Using External Device
See "Scanning Channels" in Chapter 3 for examples of scanning programs 