Advantech 19" Slim Point-of-Care Terminal POC-S196 User Manual

Page of 58
POC-S196 User Manual
4
Touchscreen (optional)
Optional modules
!
Memory: 512 MB, 1 GB, 2 GB DDR2 533/667MHz DRAM
!
HDD: 2.5" SATA HDD
!
Touchscreen: Analog resistive
Cleaning/Disinfecting
During normal use, the POC-S196 may become soiled and should therefore, be
cleaned regularly. Agents: Green tinctured soap and Enzymatic detergents
Steps:
1.
Wipe the POC-S196 with a clean cloth that has been moistened in the cleaning 
solution.
2.
Prepare agent per manufacturer’s instructions or hospital protocol.
3.
Wipe thoroughly with a clean cloth.
Table 1.1: Touchscreen Specification
Type
Analog Resistive
Resolution
Continuous
Light Transmission
75%
Controller
RS-232 interface (uses COM3)
Power Consumption
+5V@200 Ma
Software Driver
Supports Windows 2000,Windows XP
Durability (touches in a lifetime)
30 million
Note!
The Point-of-Care Terminal with the optionally installed touchscreen will 
share COM3. Once the touchscreen is installed, COM3 cannot be used 
for other purposes.
Caution! Do not immerse or rinse the POC-S196 or its peripherals. If you acci-
dentally spill liquid on the device, disconnect the unit from the power 
source. Contact your Biomed department regarding the continued safety 
of the unit before placing it back in operation.
Do not spray cleaning agent on the chassis.
Do not use disinfectants that contain phenol. Do not autoclave or clean 
the POC-S196 or its peripherals with strong aromatic, chlorinated, 
ketone, ether, or Esther solvents, sharp tools or abrasives. Never 
immerse electrical connectors in water or other liquids.