Bayer HealthCare Rapidlab 800 User Manual

Page of 699
     
       
The 800 system indicates that it is ready to analyze samples when the following
conditions exist:
 Ready screen appears
 sample door is open
 Analyze key is lit
Do not attempt to analyze samples if the system is not ready.
The system aspirates all samples except expired gas samples. Aspiration reduces
exposure to biohazardous materials and increases precision:
 by sampling the correct volume for each analysis
 by ensuring precise sample heating before sample analysis
 by reducing instrument-to-instrument and operator-to-operator variability
       
The system automatically identifies the sampling device. If the system cannot
identify the device, a message box appears prompting you to select one of the
following sample devices:
 1 mL syringe
 >2 mL syringe
 capillary
Select the device type and press Enter. Then press Analyze to begin analysis.
  
If the system detects bubbles when it aspirates a sample, the Bubbles Detected in
Sample message box appears. If the sample is not at point A or beyond, position
the sample manually to continue with analysis.
      
      
    
begin analysis after you move the sample to the measurement
module.
cancel analysis. The system starts a wash and returns to the Ready
screen.