Bayer HealthCare Rapidlab 800 User Manual
The 800 system indicates that it is ready to analyze samples when the following
conditions exist:
conditions exist:
Ready screen appears
sample door is open
Analyze key is lit
sample door is open
Analyze key is lit
Do not attempt to analyze samples if the system is not ready.
The system aspirates all samples except expired gas samples. Aspiration reduces
exposure to biohazardous materials and increases precision:
exposure to biohazardous materials and increases precision:
by sampling the correct volume for each analysis
by ensuring precise sample heating before sample analysis
by reducing instrument-to-instrument and operator-to-operator variability
by ensuring precise sample heating before sample analysis
by reducing instrument-to-instrument and operator-to-operator variability
The system automatically identifies the sampling device. If the system cannot
identify the device, a message box appears prompting you to select one of the
following sample devices:
identify the device, a message box appears prompting you to select one of the
following sample devices:
1 mL syringe
>2 mL syringe
capillary
>2 mL syringe
capillary
Select the device type and press Enter. Then press Analyze to begin analysis.
If the system detects bubbles when it aspirates a sample, the Bubbles Detected in
Sample message box appears. If the sample is not at point A or beyond, position
the sample manually to continue with analysis.
Sample message box appears. If the sample is not at point A or beyond, position
the sample manually to continue with analysis.
begin analysis after you move the sample to the measurement
module.
module.
cancel analysis. The system starts a wash and returns to the Ready
screen.
screen.