Barco DICOM THEATER MGP 15 User Manual

Page of 102
Troubleshooting the stabilization process
91
Troubleshooting the stabilization process
The projector stabilization has stopped unexpectedly
Possible causes:
No communication with projector. The projector may be 
switched off, removed, or the RS-232 cable may be 
disconnected at the projector’s or PC’s side.
No communication with optical sensor. The sensor 
connection to the PC may be disconnected.
The optical sensor is not or no longer positioned on the 
sensor patch.
The user has selected another signal input or another preset 
for the projector.
In all of these cases, you must restore any problems in the 
connections (if appropriate). Then you must run MediCal Pro and 
select the proper Preset. This should restart the stabilization.
The PC does not boot or does not recognise the keyboard 
after booting.
When a USB keyboard and the optical sensor are both connected 
to the PC’s USB bus, there may be a problem at start-up. This is 
not the case on all PCs.
A workaround is to connect the keyboard to the PS/2 connector.