Barco DICOM THEATER LITE MGP D5 User Manual

Page of 116
Using MediCal Pro
78
MGP D5 Dicom Theater Lite
Using MediCal Pro
After installation of the hardware and software, you must start 
MediCal Pro to calibrate the display(s).
Proceed as follows:
1. Start MediCal Pro. If appropriate, you can connect to MediCal 
Administrator.
2. During startup, MediCal Pro will detect the connected review 
display(s). They will be added automatically to the system 
configuration in MediCal Pro.
Note: In case the review displays were not automatically 
added to the configuration, you can add them manually, by 
running the Configuration Setup Wizard.
3. For all the devices in the configuration, check if the DPMS 
setting is turned on. Therefore, right-click on the display icon 
and select Properties... from the drop-down menu. Then click 
on Details... 
Check if the Powersave option is checked. If not, check it. 
This is necessary to use the DPMS possibilities of the imaging 
board.
4. For all the devices in the configuration, define (if necessary) 
and select a Preset.
5. After selecting the Preset, MediCal Pro starts consistency 
calibration automatically.
Please refer to the MediCal Pro User Guide for more information.