OPERATING INSTRUCTIONS 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
CONNECTING TWO POWER SUPPLIES IN PARALLEL 
 
Two power supplies may be connected in parallel to double the 
maximum load current. In this configuration the two supplies will 
provide two  0-30 volt output at up to 6 amps (heavier gauge hook-
up leads are advisable). Current equalizing resistors must be used 
as shown in Fig. 9. However, the protective current limiting 
feature will prevent damage if current is temporarily unbalanced 
during set -up. 
 
When connected in parallel and operating in the constant 
voltage mode, determine the total load current limit and preset the 
current limiting for each power supply to half the total load 
current value. Then when the load is connected, set the 
VOLTAGE controls on the two power supplies for equal voltage 
readings. This should also provide approximately equal current 
from each supply. Add the two current meter readings together for 
total load current, or connect an external ammeter in series with 
the load. 
 
If the current equalizing resistors are not well matched, it is 
preferable that the voltages be slightly unbalanced to achieve 
current balance. Be sure that the supplies are adequately 
balanced so that both remain in the CV mode. 
 
When conn ected in parallel and operating in the constant 
current mode, the  VOLTAGE controls of both supplies should 
be preset to the same value. Then when the load is connected, the 
CURRENT controls of the two supplies should be adjusted for 
approximately equal current from each unit. Be sure that both 
supplies remain in the CC mode.