Well REF 89000 User Manual

Page of 19
 
       REVOLUTION
  
 
 
 
 
                ENGLISH
 
                          45MHz ROTATIONAL IMAGING CATHETER              
                            
CAUTION: 
1.  U.S. Federal Law restricts this device to sale by or on the order of a 
 physician. 
2.  Prior to use, read this entire Instructions For Use.
 
INTENDED USE: 
The Revolution
®
 catheter is intended for the intravascular ultrasound 
examination of coronary arteries. Intravascular ultrasound imaging is indicated in 
patients who are candidates for transluminal interventional procedures. 
DESCRIPTION: 
The Revolution 45MHz rotational IVUS imaging catheter consists of two main 
assemblies; the imaging core and the catheter body. The catheter body 
comprises of three sections; distal section w/ .014" compatible F/X port, 
proximal section (single lumen), telescope section. 
The distal section and proximal (single lumen) sections comprise the "working 
length" of the catheter, the telescoping section remains outside of the guiding 
catheter. The telescoping shaft (section) allows the imaging core to be advanced 
and retracted for up to 150 mm of linear movement. The corresponding movement 
of the transducer occurs from the proximal end of the guidewire exit port to the 
proximal end of the window portion of the distal section. 
The imaging core is composed of a hi-torque, flexible, rotating drive cable with a 
distal outward looking 45MHz ultrasonic transducer. An electromechanical 
connector interface at the proximal end makes the connection to the patient 
interface module (PIM). The PIM-catheter interface consists of an integrated 
mechanical drive assembly and electrical connection. 
A flushing port with a one-way valve (Fig. 1) is used to displace the air initially 
present within the catheter. The catheter must be flushed with heparinized 
saline prior to use, as this provides the acoustic coupling media required for 
ultrasonic imaging. The one-way valve helps retain saline in the catheter 
during use. 
The catheter body has a distal guidewire lumen with a proximal exit port 
located 2 cm from the distal end (Fig 1). A radiopaque (RO) marker is 
embedded in the catheter body at 0.5 cm from the tip. In addition, an insertion 
depth indicator is located on the catheter body at 100 cm, corresponding to 
femoral insertions. 
The catheter is for use with the In-Vision Gold imaging system with software 
V5.0 or higher or the Volcano s5 and Volcano s5i imaging system. Consult 
your System Operator’s Manual. 
CONTRAINDICATIONS: 
This device is not currently indicated for use in cerebral or peripheral vessels. 
Use of IVUS Imaging Catheters is contraindicated where introduction of any 
catheter would constitute a threat to patient safety. Contraindications include: 
bateremia or sepsis, major coagulation system abnormalities, patients 
disqualified for CABG surgery, patients disqualified for PTCA, severe 
hemodynamic instability or shock, patients diagnosed with coronary artery 
spasm, and total occlusion. 
ADVERSE EFFECTS: 
Bleeding at the entry puncture site, injury to the vascular wall, thrombosis of 
the vessel, and peripheral embolization has occurred with the use of 
percutaneous intravascular catheter devices. 
 
 
WARNINGS: 
ƒ
  Use of the Revolution catheters is restricted to specialists who are familiar 
with, and have been trained to perform, the procedures for which this device 
is intended. 
ƒ
  DO NOT advance the catheter if resistance is encountered. The catheter 
should never be forcibly inserted into lumens narrower than the catheter 
body or forced through a tight stenosis. 
ƒ
  Care should be taken when utilizing devices that comprise a short monorail; 
in such instances advancement of the device distal to a deployed stent can 
result in exposure of the guidewire to the stent struts. 
ƒ
  In instances where the device has crossed a deployed stent, care should be 
taken when retracting the device to ensure that entanglement does not 
occur. Fluoroscopy should be used to monitor guidewire position with the 
respect to the imaging catheter and the stent; at no time should the imaging 
catheter be retracted if there is evidence of guidewire prolapse or if 
significant resistance to withdrawal is experienced. If either of these events 
occur, advance the imaging catheter distal of the stent and then carefully 
remove the whole system under the guidance of fluoroscopy. 
ƒ
  Care should be taken when re-advancing a guide wire after stent deployment. A 
guide wire may exit between stent struts when re-crossing a stent that is not fully 
apposed to the vessel wall. Subsequent advancement of the catheter could 
cause entanglement between the catheter and the stent. Care should be taken 
to slowly remove the catheter from a stented vessel. 
PRECAUTIONS: 
The Revolution device is a delicate scientific instrument and should be treated 
as such. Always observe the following precautions: 
ƒ
  Contents supplied STERILE using an EtO (ethylene oxide) process. Do not 
use if sterile barrier is damaged. If damage is found call your Volcano 
Corporation representative. 
ƒ
  To maintain optimal patient safety, inspect the product prior to use. Do not 
use if saline leaks from any location other than the vent port in the monorail 
section.  
ƒ
 For single use only. Do not re-use, reprocess or re-sterilize. Re-use, 
reprocessing or re-sterilization may compromise the structural integrity of 
the device and/or lead to device failure which in turn may result in patient 
injury, illness or death. 
ƒ
 Re-use, reprocessing or re-sterilization may also create a risk of 
contamination of the device and/or cause patient infection or cross-infection, 
including, but not limited to the transmission of infectious disease(s) from 
one patient to another. Contamination of the device may lead to injury, 
illness or death of the patients. 
ƒ
  The catheter has no user serviceable parts. Do not attempt to repair or to 
alter any component of the catheter assembly.  
ƒ
  Do not attempt to connect the catheter to electronic equipment other than 
the designated systems. 
ƒ
  Never attempt to attach or detach the catheter while the PIM motor is 
running. To do so may damage the connector. 
ƒ
  Avoid any sharp bends, pinching, or crushing of the catheter. 
ƒ
  Do not kink or sharply bend the catheter at any time. This can cause drive 
cable failure. An insertion angle greater than 45º is considered excessive. 
ƒ
  Turn the PIM "OFF" before withdrawing the imaging catheter. 
 
 
Fig. 1
One Way Luer
Telescoping Section 
Pullback Length 
150 mm
Proximal Shaft
109 cm
Usable Length
135 cm
Monorail
23mm
RO
Marker
3.5F
3.2F