Welch Allyn Medical Diagnostic Equipment Crib Toy LXi User Manual

Page of 130
112
Disassembly and Repair
Welch Allyn Spot Vital Signs LXi
3.
Remove the four screws from the mounting plate and save the fish paper. This 
exposes the circuit board.
4.
Remove the three Torx screws (T10) on the board (one on bottom and one on either 
side of the probe well cover) and lift the board out of the housing and orient the part 
as shown.
5.
Disconnect all connectors and replace with a new board.
To replace the temperature connector, remove the one screw holding the temperature 
board and connector. Verify the fish paper is properly mounted upon reassembly.
When replacing the SpO2/blood pressure housing, verify the orange and white twisted 
wire is pushed as far into the 
handle
 housing as possible. Failure to do so can cause the 
wires to stretch when the SpO2/blood pressure housing is inserted.