ReliOn Prime User Manual

Page of 70
44
Note:
 Dye added to control solution may 
stain surfaces or fabric. Use with care.
Step 5
Compare the result with the range printed on 
the test strip bottle you are using. Values may 
vary from bottle to bottle. Verify that the result is 
within range of the control solution level that you 
are using. If the result falls within the range on the 
bottle, the meter is working correctly. 
Step 6
Remove the test strip. The meter shuts off automatically when the test 
strip is removed.
Dispose of the test strip immediately.
06146A 
2010-10
80-120 
200-300 
Control Solution Range (mg/dL) 
(Rango de la solución de control) 
Normal:
High:
Step 4
Bring meter and test strip to the drop. Test strip will draw up the  
solution. The meter will show result in 7 seconds.