Thermometer symbol - E6. 
Temperature outside of acceptable range of 41°F - 113°F
(5°C - 45°C). You will not obtain a reading. Allow the
meter and strips to warm up or cool down slowly to 
between 50°F - 104°F (10°C - 40°C) or at least to 41°F -
113°F (5°C - 45°C).
Thermometer flashing. 
The meter and test strips are outside the optimum 
temperature range of 50°F - 104°F (10°C - 40°C). Make
sure your meter and strips are in an environment with a 
temperature between 50°F - 104°F (10°C - 40°C). Treat
test result obtained where the thermometer is flashing 
with caution. The flashing thermometer will be displayed 
on all screens during glucose testing until the meter has 
reached acceptable temperature. Readings obtained with 
the flashing thermometer will show a thermometer when 
results are recalled in memory.