Omega DPF500 Series User Manual

Page of 18
7
4.1 Explanation of Display Prompts (continued)
oUt HI: (out high) Sets the high setting for the 4-20 mA analog output.  Key in the high rate value at which the unit 
will output 20 mA.  Press the ↑ key to increment each digit.  Press the ← key to step to the next digit to the left.  
Press the E key to enter the displayed out hi value.
pUlscale: (pulse out scaling) This allows the unit to output a pulse for each least signifi cant total count divided by 
the selected divider.  The pulse out can be divided by 1 (
d 1), 10 (d 10), 100 (d 100), or turned off (off).   With 
the divider set at 1, the unit will give a pulse out for every increment of the LSD displayed. 
Note: For maximum battery life, turn the pulse output off when pulse output is not used.  
Selecting the proper pulse output divider may be needed so that the pulse output 
does not exceed the maximum rate of the pulse output. If the pulse output 
pulses too quickly a fl ashing display will result.  Pressing the “M” key will result in a display of an error mes-
sage “E PULSE”. Press the “E” key to return to the run mode.
P uuidth: (pulse width) Sets the pulse width of the pulse output.  Selections are: 0.5 (1Hz), 0.25 (2Hz),0.125 (4Hz) 
or 0.0625 (8Hz).  This menu item is skipped if 
pUlscale is turned off.
loC Code: (lock code) Sets the 5 digit lock code to be entered when the unit prompts ent Code.  This allows the user 
to gain access to the menu when the unit is locked.  Press the ↑ key to increment each digit.  Press the ← key 
to step to the next digit to the left.  Press the E key to enter the displayed code.  
 
Record this number for later use! 
LOCK CODE: ____________________
loC UnIt: (lock unit) Sets the panel lock ON or OFF.  Press the ↑ key to select On or OFF.  Press the E key to enter 
the displayed selection.