Analog Measurements - Screens and Control Fields
Scope To
Scope To
Scope To
This field selects the signal source for the oscilloscope. This allows you
to bypass certain sections of the AF analyzer’s circuitry when viewing
and measuring a signal. It also allows you to select measurement paths
that include additional gain stages, improving the oscilloscope’s
resolution when measuring low-level signals.
to bypass certain sections of the AF analyzer’s circuitry when viewing
and measuring a signal. It also allows you to select measurement paths
that include additional gain stages, improving the oscilloscope’s
resolution when measuring low-level signals.
•
Input
looks at the unfiltered signal directly from the input.
•
Filters
looks at the signal after filtering through Filter 1 and
Filter 2.
•
De-Emp
looks at the signal after filtering through Filter 1 and
Filter 2, and after 750
µ
s de-emphasis (if the
De-emphasis
field is
set to
750 us
.)
•
Notch
looks at the signal after filtering through Filter 1 and Filter
2, after de-emphasis (if used), and after the notch filter.
Screens on which this field is displayed
AF ANALYZER
Main Menu