Maintenance
PERFORMANCE TEST
6
6-7
6-5. 
AC Voltage Test (Option -09 Only)
The following procedure may be used to verify the accuracy of the VAC function:
1.
 
Ensure the 8842A is on and warmed up for at least 1 hour.
2.
 
Select the VAC function and the slow (S) reading rate.
3.
 
Connect the AC Calibrator to provide a voltage input to the HI and LO INPUT
terminals.
4.
 
(Low- and Mid-Frequency Test.) For each step in Table 6-3, select the indicated
range, set the AC Calibrator for the specified input, and verify that the displayed
reading is within the limits shown for each reading rate.
NOTE
This procedure tests the extremes of each range. You may shorten the
procedure by testing only the "quick test points" indicated in Table 6-3 with
asterisks.
5.
 
(High-Frequency Test.) For each step in Table 6-4, select the indicated range, set the
AC Calibrator for the specified input, and verify that the displayed reading is within
the limits shown for each reading rate.
NOTE
This procedure tests the extremes of each range. You may shorten the
procedure by testing only the "quick test points" indicated in Table 6-4 with
asterisks.
6.
 
Set the AC calibrator to standby and disconnect it from the 8842A.
Table 6-3. Low- and Mid-Frequency AC Voltage Test
Input
Display Readings
Step
Number
Range
Voltage
Frequency
Error in
Counts
Minimum
Maximum
1
2V
0.01000V
200 Hz
181
0.00819 VAC
0.01181 VAC
2
2V
0.10000V
200 Hz
87
.09913 VAC
.10087 VAC
3*
2V
0.30000V
200 Hz
101
0.29899 VAC
0.30101 VAC
4
2V
1.00000V
200 Hz
150
0.99850 VAC
1.00150 VAC
5*
2V
1.90000V
200 Hz
213
1.89787 VAC
1.90213 VAC
6
2V
0.10000V
20 Hz
220
0.09780 VAC
0.10220 VAC
7
2V
1.90000V
20 Hz
2380
1.87620 VAC
1.92380 VAC
8
2V
.10000V
45 Hz
135
0.09865 VAC
0.10135 VAC
9
2V
1.90000V
45 Hz
765
1.89235 VAC
1.90765 VAC
10*
200 mV
0.001000V
100 Hz
204
00.796 mVAC
190.252 mVAC
11
200 mV
0.190000V
20 kHz
252
189.748 mVAC
190.252 MvAC
* Quick Test points