Fluke Portable Multimedia Player 8842a User Manual

Page of 269
8842A
Instruction Manual
6-20
readings. (In remote calibration, the averaged value can be stored in the controller.)
Record the value.
4.
 
Select the "100
" output from the 5450A, and measure this value as in step 3.
5.
 
Find and record the numerical difference between the values calculated in steps 3 and
4. This value will be used as the variable  input for calibrating the 200
 range in 2-
wire ohms.
6.
 
Repeat steps 4 and 5 using the "1 k
" output from the 5450A; find and record the
numerical difference between this and the "SHORT" measured in step 3. This value
will be used as the variable input for calibrating the 2 k
 range in 2-wire ohms.
7.
 
Press the UUT’s 2 WIRE k
 button. This selects the Offset and Gain calibration
procedure for the 2-wire ohms function and prompts for zero input. Select the
"SHORT" from the 5450A, and calibrate all the zeros by pressing STORE.
8.
 
Select the "100
" output from the 5450A and calibrate the high point for the 200
range, entering the value computed in step 5 as a variable input.
9.
 
Select the "1 k
" output from the 5450A and calibrate the high point for the 2 k
range, entering the value computed in step 6 as a variable input.
10.
 
Calibrate the remaining ranges (steps D-G of Table 6-10) using the 5450A outputs.
11.
 
Recalibrate the low point for each 2-wire ohms range using a shorting link (Pomona
MDP-S-0 or equivalent) across the UUT’s HI and LO INPUT terminals.
12.
 
Exit the calibration mode by pressing the CAL ENABLE switch.
13.
 
Using the same configuration shown in Figure 6-4, verify that the UUT measures the
same value (within 1 digit) in 2-wire ohms (using the offset feature to correct for
5450A floor error) as in 4-wire  ohms. If the readings differ by more than 1 digit,
reenable the calibration mode and repeat steps 2 through 8.
NOTE
Only 4-wire ohm calibration is allowed in the 20
 range.
14.
 
Cover the CAL ENABLE switch with a calibration certification sticker.
f6-04.wmf
Figure 6-4. Optimizing Use of the 5450A
6-23.  Remote Calibration
If the IEEE-488 Interface is installed, the 8842A can be calibrated under remote control.
Remote calibration is very similar to local (front-panel controlled) calibration. Table 6-14
shows the remote commands which correspond to the front panel features.To facilitate
remote calibration, there are some differences from local calibration: