Campbell Manufacturing TEMPERATURE PROBE 109SS User Manual

Page of 18
Model 109SS Temperature Probe 
The resistance of the thermistor is: 
=
1
900
,
24
V
V
R
EX
t
 
The Steinhart and Hart equation is used to calculate temperature from 
Resistance: 
3
))
(ln(
)
ln(
1
T
T
K
R
C
R
B
A
T
+
+
=
 
Where T
K
 is the temperature in Kelvin.  The Steinhart and Hart coefficients 
used in the Therm109 instruction are: 
A = 1.129241x10
-3
 
B = 2.341077x10
-4
 
C = 8.775468x10
-8 
7.  Maintenance and Calibration 
The 109SS Probe requires minimal maintenance.  Periodically check cabling 
for proper connections, signs of damage, and possible moisture intrusion.  For 
all factory repairs and recalibrations, customers must get a returned materials 
authorization (RMA).  Customers must also properly fill out a “Declaration of 
Hazardous Material and Decontamination” form, and comply with the 
requirements specified in it.  Refer to the “Warranty and Assistance” page for 
more information. 
8.  Troubleshooting 
Symptom:  Temperature is NAN, -INF, -9999, -273 
Verify the red wire is connected to the correct Single-Ended analog input 
channel as specified by the measurement instruction, the black wire is 
connected to the switched excitation channel as specified by the measurement 
instruction, and the purple wire is connected to datalogger ground.  
Symptom:  Incorrect Temperature 
Verify the multiplier and offset parameters are correct for the desired units 
(Section 5).  Check the cable for signs of damage and possible moisture 
intrusion. 
For all factory repairs, customers must get an RMA.  Customers 
must also properly fill out a “Declaration of Hazardous Material 
and Decontamination” form and comply with the requirements 
specified in it.  Refer to the “Warranty and Assistance” page for 
more information. 
NOTE 
11