Model 109SS Temperature Probe 
2.  Accuracy 
The overall probe accuracy is a combination of the thermistor's interchangeability 
specification and the accuracy of the bridge resistor.  The Steinhart and Hart 
equation used to calculate temperature has a negligible error (Figure 2-1).  In a 
"worst case" the errors add to an accuracy of ±0.6
°C over the range of -40° to 70°C 
and 
±0.49°C over the range of -20°C to 70°C.  The major error component is the 
interchangeability specification (tolerance) of the thermistor.  The bridge resistor 
has a 0.1% tolerance with a 10 ppm temperature coefficient.  Figure 2-2 shows the 
possible worst case probe and measurement errors.   
Steinhart & Hart  - Tabulated values
-0.005
0
0.005
0.01
0.015
0.02
0.025
0.03
-50
-40
-30
-20
-10
0
10
20
30
40
50
60
70
Temperature Degrees C
Error Degrees C
 
FIGURE 2-1.  Steinhart and Hart 
 
FIGURE 2-2.  Possible Errors